Test Code CLFA; SQ: CRYCSF Cryptococcus Antigen Screen with Titer, Spinal Fluid
Useful For
Aiding in the diagnosis of cryptococcosis
This test should not be performed as a screening procedure for the general population.
This test should not be used as a test of cure or to guide treatment decisions.
Disease States
- Cryptococcosis
Reflex Tests
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
CLFAT | Cryptococcus Ag Titer, LFA, CSF | Yes | No |
Testing Algorithm
If result is positive, Cryptococcus titer will be performed at an additional charge.
For more information see Meningitis/Encephalitis Panel Algorithm
Reporting Name
Cryptococcus Ag Screen w/Titer, CSFSpecimen Type
CSFSpecimen Required
Container/Tube: Sterile vial
Specimen Volume: 1 mL
Specimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
CSF | Refrigerated (preferred) | 14 days | |
Frozen | 14 days |
Reject Due To
Gross hemolysis | Reject |
Clinical Information
Cryptococcosis is an invasive fungal infection caused by Cryptococcus neoformans or Cryptococcus gattii. C neoformans has been isolated from several sites in nature, particularly weathered pigeon droppings. C gattii was previously only associated with tropical and subtropical regions. More recently, however, this organism has been found to be endemic in British Columbia and the Pacific Northwestern United States and is associated with several different tree species.
Infection is usually acquired via the pulmonary route. Patients are often unaware of any exposure history. Approximately half of the patients with symptomatic disease have a predisposing immunosuppressive condition such as AIDS, steroid therapy, lymphoma, or sarcoidosis. Symptoms may include fever, headache, dizziness, ataxia, somnolence, and cough. While the majority of C neoformans infections occur in immunocompromised patient populations, C gattii has a higher predilection for infection of healthy individuals.(1,2)
In addition to the lungs, cryptococcal infections frequently involve the central nervous system (CNS), particularly in patients infected with HIV. Mortality among patients with CNS cryptococcosis may approach 25% despite antibiotic therapy. Untreated CNS cryptococcosis is invariably fatal. Disseminated disease may affect any organ system and usually occurs in immunosuppressed individuals.
Reference Values
Negative
Reference values apply to all ages.
Interpretation
The presence of cryptococcal antigen in any body fluid (serum or cerebrospinal fluid [CSF]) is indicative of cryptococcosis. Specimens that are positive by the lateral flow assay (LFA) screen are automatically repeated by the same method utilizing dilutions to generate a titer value. CSF specimens submitted for initial diagnosis that test positive by LFA should also be submitted for routine fungal culture. Culture can aid in differentiating between the 2 common Cryptococcus species causing disease (Cryptococcus neoformans and Cryptococcus gattii) and can be used for antifungal susceptibility testing, if necessary. CSF specimens submitted to monitor antigen levels during treatment do not need to be cultured.
Disseminated infection is usually accompanied by a positive serum test.
Higher Cryptococcus antigen titers appear to correlate with more severe infections. Declining titers may indicate regression of infection. However, monitoring titers to cryptococcal antigen should not be used as a test of cure or to guide treatment decisions. Low-level titers may persist for extended periods of time following appropriate therapy and the resolution of infection.(3)
Method Description
The Cryptococcus antigen (CrAg) lateral flow assay is a sandwich immunochromatographic assay. Specimens and diluent are added to a test tube and the lateral flow device is added. The test uses specimen wicking to capture gold-conjugated, anti-cryptococcal antigen monoclonal antibodies and gold-conjugated control antibodies deposited on the test membrane. If cryptococcal antigen is present in the specimen, it binds to the gold-conjugated, anti-cryptococcal antigen antibodies. This complex wicks up the membrane and interacts with the test line, which has immobilized anti-cryptococcal antigen monoclonal antibodies. The antigen-antibody complex forms a sandwich at the test line causing a visible line to form. A valid test shows a visible line at the control line. Positive test results create 2 lines (control and specimen), while negative results form only the control line.(Package insert: CrAg Lateral Flow Assay. IMMY; Rev 06/27/2019)
Day(s) Performed
Monday through Sunday
Performing Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
87899-Cryptococcus screen
87899-Cryptococcus titer (if appropriate)
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
CLFA | Cryptococcus Ag Screen w/Titer, CSF | 29896-8 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
62074 | Cryptococcus Ag Screen w/Titer, CSF | 29896-8 |
Method Name
Lateral Flow Assay (LFA)
Secondary ID
62074Forms
If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.