Test Code CSMTU Controlled Substance Monitoring Targeted Profile, 17 Drug Classes, Mass Spectrometry, Random, Urine
Ordering Guidance
This test does not screen for drug classes other than those listed in Reference Values.
Specimen Required
Supplies: Urine Tubes, 10 mL (T068)
Collection Container/Tube: Plastic urine container
Submission Container/Tube: Plastic, 10 mL tube
Specimen Volume: 5 mL
Collection Instructions:
1. Collect a random urine specimen.
2. Submit 5 mL in 1 plastic bottle.
3. No preservative
Additional Information:
1. No specimen substitutions.
2. Submitting less than 5 mL may compromise the ability to perform all necessary testing.
3. STAT requests are not accepted for this test.
Secondary ID
615292Useful For
Detecting drug use involving stimulants, benzodiazepines, and opioids
This test is not intended for use in employment-related testing.
Profile Information
Test ID | Reporting Name | Available Separately | Always Performed |
---|---|---|---|
LPCM | List Patient's Current Medications | No | Yes |
ADULT | Adulterants Survey, U | Yes | Yes |
TOPSU | Targeted Opioid Screen, U | Yes, (order TOSU) | Yes |
TABSU | Targeted Benzodiazepine Screen, U | Yes, (order TBSU) | Yes |
TSTIM | Targeted Stimulant Screen, U | Yes, (order TSPU) | Yes |
Testing Algorithm
Testing begins with an adulterant survey. If the sample is found to be adulterated, testing will end, and the remaining tests will be canceled.
If the specimen is normal or only diluted, testing will proceed.
Method Name
ADULT: Spectrophotometry
TOPSU, TABSU, TSTIM: Liquid Chromatography Tandem Mass Spectrometry, High-Resolution Accurate Mass (LC-MS/MS HRAM)
Reporting Name
CSM Targeted Drug Profile,17,HRMS,USpecimen Type
UrineSpecimen Minimum Volume
2 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Urine | Refrigerated (preferred) | 14 days | |
Frozen | 14 days |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Clinical Information
The targeted opioid, benzodiazepine, and stimulant screen portions are performed by liquid chromatography tandem mass spectrometry, high-resolution accurate mass (LC-MS/MS HRAM) and are completed for all opioids, benzodiazepines, and stimulants.
Opioids are a large class of medications commonly used to relieve acute and chronic pain or help manage opioid abuse and dependence. Medications that fall into this class include buprenorphine, codeine, fentanyl, hydrocodone, hydromorphone, methadone, morphine, oxycodone, oxymorphone, tapentadol, tramadol, and others. Opioids work by binding to the opioid receptors that are found in the brain, spinal cord, gastrointestinal tract, and other organs. Common side effects of opioids include drowsiness, confusion, nausea, constipation, and, in severe cases, respiratory depression. These are dose dependant and vary with tolerance. These medications can also produce physical and psychological dependence and have a high risk for abuse and diversion, which is one of the main reasons many professional practice guidelines recommend compliance testing in patients prescribed these medications.
Opioids are readily absorbed from the gastrointestinal tract, nasal mucosa, lungs, and after subcutaneous or intermuscular injection. Opioids are primarily excreted from the kidney in both free and conjugated forms. This assay does not hydrolyze the urine sample and looks for both parent drugs and metabolites (including glucuronide forms). The detection window for most opioids in urine is approximately 1 to 3 days with longer detection times for some compounds (eg, methadone).
Benzodiazepines represent a large family of medications used to treat a wide range of disorders from anxiety to seizures and are also used in pain management. With a high risk for abuse and diversion, professional practice guidelines recommend compliance monitoring for these medications using urine drug tests. However, traditional benzodiazepine immunoassays suffer from a lack of cross-reactivity with all the benzodiazepines, so many compliant patients taking either clonazepam (Klonopin) or lorazepam (Ativan) may screen negative by immunoassay but are positive when confirmatory testing is done. The new targeted benzodiazepine screening test provides a more sensitive and specific test to check for compliance to all the commonly prescribed benzodiazepines and looks for both parent drug and metabolites in the urine.
Stimulants are sympathomimetic amines that stimulate the central nervous system activity and, in part, suppress the appetite. Amphetamine and methamphetamine are also prescription drugs used in the treatment of narcolepsy and attention-deficit disorder/attention-deficit hyperactivity disorder (ADHD). Methylphenidate is another stimulant used to treat ADHD. Phentermine is indicated for the management of obesity. All other amphetamines (eg, methylenedioxymethamphetamine: MDMA) are Drug Enforcement Administration-scheduled Class I compounds. Due to their stimulant effects, the drugs are commonly sold illicitly and abused. Physiological symptoms associated with very high amounts of ingested amphetamine or methamphetamine include elevated blood pressure, dilated pupils, hyperthermia, convulsions, and acute amphetamine psychosis.
This test is intended to be used in a setting where the test results can be used to make a definitive diagnosis.
Reference Values
ADULTERANT SURVEY:
Cutoff concentrations
Oxidants: 200 mg/L
Nitrites: 500 mg/L
TARGETED OPIOID SCREEN:
Not detected
Cutoff concentrations:
Codeine: 25 ng/mL
Codeine-6-beta-glucuronide: 100 ng/mL
Morphine: 25 ng/mL
Morphine-6-beta-glucuronide: 100 ng/mL
6-Monoacetylmorphine: 25 ng/mL
Hydrocodone: 25 ng/mL
Norhydrocodone: 25 ng/mL
Dihydrocodeine: 25 ng/mL
Hydromorphone: 25 ng/mL
Hydromorphone-3-beta-glucuronide: 100 ng/mL
Oxycodone: 25 ng/mL
Noroxycodone: 25 ng/mL
Oxymorphone: 25 ng/mL
Oxymorphone-3-beta-glucuronide: 100 ng/mL
Noroxymorphone: 25 ng/mL
Fentanyl: 2 ng/mL
Norfentanyl: 2 ng/mL
Meperidine: 25 ng/mL
Normeperidine: 25 ng/mL
Naloxone: 25 ng/mL
Naloxone-3-beta-glucuronide: 100 ng/mL
Methadone: 25 ng/mL
2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP): 25 ng/mL
Propoxyphene: 25 ng/mL
Norpropoxyphene: 25 ng/mL
Tramadol: 25 ng/mL
O-desmethyltramadol: 25 ng/mL
Tapentadol: 25 ng/mL
N-desmethyltapentadol: 50 ng/mL
Tapentadol-beta-glucuronide: 100 ng/mL
Buprenorphine: 5 ng/mL
Norbuprenorphine: 5 ng/mL
Norbuprenorphine glucuronide: 20 ng/mL
TARGETED BENZODIAZEPINE SCREEN:
Not detected
Cutoff concentrations:
Alprazolam: 10 ng/mL
Alpha-hydroxyalprazolam: 10 ng/mL
Alpha-hydroxyalprazolam glucuronide: 50 ng/mL
Chlordiazepoxide: 10 ng/mL
Clobazam: 10 ng/mL
N-desmethylclobazam: 200 ng/mL
Clonazepam: 10 ng/mL
7-Aminoclonazepam: 10 ng/mL
Diazepam: 10 ng/mL
Nordiazepam: 10 ng/mL
Flunitrazepam: 10 ng/mL
7-Aminoflunitrazepam: 10 ng/mL
Flurazepam: 10 ng/mL
2-Hydroxy ethyl flurazepam: 10 ng/mL
Lorazepam: 10 ng/mL
Lorazepam glucuronide: 50 ng/mL
Midazolam: 10 ng/mL
Alpha-hydroxymidazolam: 10 ng/mL
Oxazepam: 10 ng/mL
Oxazepam glucuronide: 50 ng/mL
Prazepam: 10 ng/mL
Temazepam: 10 ng/mL
Temazepam glucuronide: 50 ng/mL
Triazolam: 10 ng/mL
Alpha-hydroxytriazolam: 10 ng/mL
Zolpidem: 10 ng/mL
Zolpidem phenyl-4-carboxylic acid: 10 ng/mL
TARGETED STIMULANT SCREEN:
Not detected
Cutoff concentrations:
Methamphetamine: 100 ng/mL
Amphetamine: 100 ng/mL
3,4-Methylenedioxymethamphetamine (MDMA): 100 ng/mL
3,4-Methylenedioxy-N-ethylamphetamine (MDEA): 100 ng/mL
3,4-Methylenedioxyamphetamine (MDA): 100 ng/mL
Ephedrine: 100 ng/mL
Pseudoephedrine: 100 ng/mL
Phentermine: 100 ng/mL
Phencyclidine (PCP): 20 ng/mL
Methylphenidate: 20 ng/mL
Ritalinic acid: 100 ng/mL
Interpretation
A positive result derived by this testing indicates that the patient has used one of the drugs detected by these techniques in the recent past.
For information about drug testing, including estimated detection times and Result Interpretations, see Controlled Substance Monitoring on MayoClinicLabs.com.
Method Description
Adulterant:
All results are measured using spectrophotometry at wavelengths specified by the reagent manufacturer. The use of a refractometer may also be used in the specific gravity measurement.(Package inserts: Specimen Validity Test Creatinine. Roche Diagnostics; V3.0, 08/2015; Specimen Validity Test Nitrite. Roche Diagnostics; V3.0, 08/2018, Specimen Validity Test Oxidant. Roche Diagnostics; V 3.0, 08/2018; Specimen Validity Test pH Roche Diagnostics; V3.0, 02/2019, Specimen Validity Test Specific Gravity. Roche Diagnostics; V4.0, 08/2022)
Targeted Screening Panels for opioids, benzodiazepines, and stimulants:
The urine sample is diluted with internal standard and clinical laboratory reagent water and then analyzed by liquid chromatography tandem mass spectrometry using a high-resolution accurate mass orbitrap detector.(Unpublished Mayo method)
Day(s) Performed
Monday through Saturday
Performing Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
80347
80364
80326
G0482 (if appropriate)
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
CSMTU | CSM Targeted Drug Profile,17,HRMS,U | 69739-1 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
LPCM | List Patient's Current Medications | 66423-5 |
610273 | Methamphetamine | 19554-5 |
20606 | Creatinine, U | 2161-8 |
604871 | Alprazolam | 94116-1 |
42323 | Codeine | 19411-8 |
42324 | Codeine-6-beta-glucuronide | 89310-7 |
604867 | Alpha-Hydroxyalprazolam | 19325-0 |
22312 | Specific Gravity | 5810-7 |
610274 | Amphetamine | 19343-3 |
610275 | 3,4-methylenedioxymethamphetamine (MDMA) | 19568-5 |
23509 | pH | 2756-5 |
604891 | Alpha-Hydroxyalprazolam Glucuronide | 94115-3 |
42325 | Morphine | 19597-4 |
42326 | Morphine-6-beta-glucuronide | 89308-1 |
604872 | Chlordiazepoxide | 19385-4 |
23511 | Oxidants | 58714-7 |
610276 | 3,4-methylenedioxy-N-ethylamphetamine (MDEA) | 59844-1 |
610277 | 3,4-methylenedioxyamphetamine (MDA) | 19565-1 |
23510 | Nitrites | 32710-6 |
604889 | Clobazam | 94114-6 |
42327 | 6-monoacetylmorphine | 19321-9 |
42328 | Hydrocodone | 19482-9 |
604890 | N-Desmethylclobazam | 94113-8 |
30914 | Comment | 48767-8 |
610278 | Ephedrine | 99108-3 |
610279 | Pseudoephedrine | 99109-1 |
604873 | Clonazepam | 19399-5 |
42329 | Norhydrocodone | 89304-0 |
42330 | Dihydrocodeine | 19446-4 |
604267 | 7-aminoclonazepam | 94112-0 |
610280 | Phentermine | 19674-1 |
610281 | Phencyclidine (PCP) | 19659-2 |
604874 | Diazepam | 19443-1 |
42331 | Hydromorphone | 19486-0 |
42332 | Hydromorphone-3-beta-glucuronide | 89309-9 |
604880 | Nordiazepam | 19624-6 |
610282 | Methylphenidate | 19577-6 |
610283 | Ritalinic acid | 99110-9 |
604875 | Flunitrazepam | 19466-2 |
42333 | Oxycodone | 19642-8 |
42334 | Noroxycodone | 89303-2 |
604866 | 7-aminoflunitrazepam | 94111-2 |
610284 | Stimulant Interpretation | 54247-2 |
604876 | Flurazepam | 19474-6 |
42335 | Oxymorphone | 19646-9 |
42336 | Oxymorphone-3-beta-glucuronide | 89301-6 |
604868 | 2-Hydroxy Ethyl Flurazepam | 94110-4 |
604877 | Lorazepam | 19520-6 |
42337 | Noroxymorphone | 89302-4 |
42338 | Fentanyl | 59673-4 |
604878 | Lorazepam Glucuronide | 94109-6 |
604879 | Midazolam | 19585-9 |
42339 | Norfentanyl | 43199-9 |
42340 | Meperidine | 19532-1 |
604869 | Alpha-Hydroxy Midazolam | 94108-8 |
604881 | Oxazepam | 19638-6 |
42341 | Normeperidine | 27920-8 |
42342 | Naloxone | 42618-9 |
604882 | Oxazepam Glucuronide | 94107-0 |
604883 | Prazepam | 19678-2 |
42343 | Naloxone-3-beta-glucuronide | 89307-3 |
42344 | Methadone | 19550-3 |
604884 | Temazepam | 19698-0 |
604885 | Temazepam Glucuronide | 94106-2 |
42345 | EDDP | 93495-0 |
42346 | Propoxyphene | 19429-0 |
604886 | Triazolam | 19714-5 |
604870 | Alpha-Hydroxy Triazolam | 94105-4 |
42347 | Norpropoxyphene | 19632-9 |
42348 | Tramadol | 19710-3 |
604887 | Zolpidem | 94104-7 |
604888 | Zolpidem Phenyl-4-Carboxylic acid | 94103-9 |
42349 | O-desmethyltramadol | 86453-8 |
42350 | Tapentadol | 72485-6 |
604949 | Benzodiazepine Interpretation | 69050-3 |
42351 | N-desmethyltapentadol | 89306-5 |
42352 | Tapentadol-beta-glucuronide | 89300-8 |
42353 | Buprenorphine | 93494-3 |
42354 | Norbuprenorphine | 82371-6 |
42355 | Norbuprenorphine glucuronide | 89305-7 |
65059 | Opioid Interpretation | 69050-3 |
Forms
If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.