Test Code IRON Iron, Serum
Specimen Required
Only orderable as part of profile. For more information see SFEC / Iron and Total Iron-Binding Capacity, Serum
Patient Preparation:
1. Fasting (12 hours)
2. For 24 hours before collection, patient should not take iron-containing supplements.
Supplies: Sarstedt Aliquot Tube 5 mL (T914)
Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 1 mL
Collection Instructions:
1. Draw blood before noon (preferred).
2. Within 2 hours of collection, serum gel tubes should be centrifuged.
3. Within 2 hours of collection, red-top tubes should be centrifuged, and the serum aliquoted into a plastic vial.
Secondary ID
2502Useful For
Aiding in the evaluation of iron deficiency and iron overload diseases in combination with total iron binding capacity and percent saturation
Assessment of acute iron poisoning
Method Name
Only orderable as part of profile. For more information see SFEC / Iron and Total Iron-Binding Capacity, Serum
Colorimetric Assay
Reporting Name
IronSpecimen Type
SerumSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 21 days | |
Frozen | 365 days |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | OK |
Gross icterus | OK |
Clinical Information
Iron (Fe) is involved in the function of all cells. Systemic iron homeostasis is maintained by the tight regulation of communication between cells that absorb iron from the diet (duodenal enterocytes), cells that consume iron (mainly erythroid precursors), and cells that store iron (hepatocyte and tissue macrophages). Once ingested iron is absorbed and temporarily stored in the mucosal cells within ferritin. Ferritin provides a soluble protein shell to encapsulate a complex of insoluble ferric hydroxide and ferric phosphate. Iron is released into the blood and transported as Fe (III)-transferrin.(1)
The concentration of iron varies widely, both in normal healthy persons, and in various clinical disorders. The biologic variation of iron is notable in normal healthy persons and in various clinical disorders owing to both diurnal variation and post-prandial effects. Normally, intra-individual serum iron levels undergo significant within day and day-to-day variation. The intraindividual day-to-day variation of iron is approximately 25% to 30%.(1) Serum iron concentration is frequently highest in the morning and declines progressively during the day, to reach a low point near midnight. There are no definitive studies that suggest fasting from food is required; however, serum iron is commonly drawn in the fasting state. Drawing blood three hours after consuming oral iron supplements has been shown to significantly increase serum iron concentration 3 to 5-fold and therefore should be avoided for 24 hours prior to collection.(2)
Reference Values
Only orderable as part of profile. For more information see SFEC / Iron and Total Iron-Binding Capacity, Serum
Males: 50-150 mcg/dL
Females: 35-145 mcg/dL
Interpretation
Serum iron is elevated in iron overload conditions including hemochromatosis. Additional causes include oral or parenteral intake of medicinal iron, acute hepatitis, and chronic liver failure.(1)
Serum iron is decreased in iron deficiency, iron deficiency anemia, and anemia of chronic disease.(1)
Method Description
Under acidic conditions, iron is liberated from transferrin. Lipemic samples are clarified by the detergent. Ascorbate reduces the released Fe(3+) ions to Fe(2+) ions which then react with FerroZine to form a colored complex. The color intensity is directly proportional to the iron concentration and can be measured photometrically.(Package insert: IRON2. Roche Diagnostics; V9.0, 09/2019)
Day(s) Performed
Monday through Sunday
Performing Laboratory
Mayo Clinic Laboratories in RochesterTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
83540-Iron
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
IRON | Iron | 2498-4 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
IRON | Iron | 2498-4 |