Sign in →

Test Code RA1U Alpha-1-Microglobulin, Random, Urine


Specimen Required


Container/Tube: Plastic, 5-mL tube

Specimen Volume: 5 mL

Collection Instructions:

1. Collect a random urine specimen.

2. No preservative.


Secondary ID

610364

Useful For

Assessment of renal tubular injury or dysfunction using random urine specimens

 

Screening for tubular abnormalities

 

Detecting chronic asymptomatic renal tubular dysfunction (2)

Profile Information

Test ID Reporting Name Available Separately Always Performed
AIMR Alpha-1-Microglobulin, Random, U No Yes
CRETR Creatinine, Random, U Yes, (Order RCTUR) Yes

Method Name

AIMR: Immunonephelometry

CRETR: Enzymatic Colorimetric Assay

Reporting Name

Alpha-1-Microglobulin, Random, U

Specimen Type

Urine

Specimen Minimum Volume

1 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Urine Refrigerated (preferred) 7 days
  Frozen  7 days

Reject Due To

All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.
 

Clinical Information

Alpha-1-microglobulin is a low-molecular-weight protein of 26 kDa and a member of the lipocalin protein superfamily.(1) It is synthesized in the liver, freely filtered by glomeruli, and reabsorbed by renal proximal tubules cells where it is catabolized.(1) Due to extensive tubular reabsorption, under normal conditions very little filtered alpha-1-microglobulin appears in the final excreted urine. Therefore, an increase in the urinary concentration of alpha-1-microglobulin indicates proximal tubule injury and/or impaired proximal tubular function.

 

Elevated excretion rates can indicate tubular damage associated with renal tubulointerstitial nephritis or tubular toxicity from heavy metal or nephrotoxic drug exposure. Glomerulonephropathies and renal vasculopathies also are often associated with coexisting tubular injury and so may result in elevated urinary alpha-1-microglobulin excretion. Elevated alpha-1-microglobulin in patients with urinary tract infections may indicate renal involvement (pyelonephritis). Measurement of urinary excretion of retinol-binding protein, another low-molecular-weight protein, is an alternative to the measurement of alpha-1-microglobulin. To date, there are no convincing studies to indicate that one test has better clinical utility than the other.

 

Urinary excretion of alpha-1-microglobulin can be determined from either a 24-hour collection or from a random urine collection. The 24-hour collection is traditionally considered the gold standard. For random or spot collections, the concentration of alpha-1-microglobulin is divided by the urinary creatinine concentration. This corrected value adjusts alpha-1-microglobulin for variabilities in urine concentration.

Reference Values

≥18 years: <35 mg/g creatinine

Reference values have not been established for patients who are less than 18 years of age.

Interpretation

Alpha-1-microglobulin above the reference values may indicate a proximal tubular dysfunction. As suggested in the literature, 7 mg/g creatinine is an upper reference limit for pediatric patients of 1 month to 15 years of age.(3)

Method Description

Alpha-1-Microglobulin:

In an immunochemical reaction, alpha-1-microglobulin present in the urine sample forms immune complexes with anti-alpha-1-microglobulin-specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of alpha-1-microglobulin in the sample. The result is evaluated by comparison with a standard of known concentration.(Package insert: N Alpha-1-microglobulin. Siemens; V5, 05/2018)

 

Creatinine:

Creatinine is performed by the enzymatic method, which is based on the determination of sarcosine from creatinine with the aid of creatininase, creatinase, and sarcosine oxidase. The liberated hydrogen peroxide is measured via a modified Trinder reaction using a colorimetric indicator. Optimization of the buffer system and the colorimetric indicator enables the creatinine concentration to be quantified both precisely and specifically.(Package insert: Creatinine plus ver 2. Roche Diagnostics; V15.0, 03/2019)

Day(s) Performed

Varies

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

83883

LOINC Code Information

Test ID Test Order Name Order LOINC Value
RA1U Alpha-1-Microglobulin, Random, U 48415-4

 

Result ID Test Result Name Result LOINC Value
A1M_U Alpha-1-Microglobulin, Random, U 46723-3
CRETR Creatinine, Random, U 2161-8
A1M_R A1M/Creat Ratio 48415-4

Forms

If not ordering electronically, complete, print, and send a Renal Diagnostics Test Request (T830) with the specimen.